Basics of spectrophotometry and optical measurements

Transmittance, absorbance and the Beer–Lambert law; spectrophotometer sources, monochromators, gratings, detectors, single/double beam and sources of error.

Drafted with Aria, reviewed by the AiCanCode.org team. Spotted an error? Use Give Feedback at the bottom of the page.

Why it matters

Spectrophotometers measure how much light a sample absorbs at each wavelength. They are the workhorse analysers of water-quality labs, pharmaceutical QC, clinical chemistry and process plants (on-line colour and concentration monitors), and the same building blocks — sources, monochromators, detectors and ratio measurement — appear in optical power meters, spectrum analysers and gas analysers. An instrumentation engineer must be able to turn a transmittance reading into a concentration and judge when that reading can be trusted.

Key ideas

Transmittance and absorbance. A beam of radiant power P₀ enters a sample and P leaves it. Transmittance is T = P/P₀ (often quoted as %T). Absorbance is A = −log₁₀T = log₁₀(P₀/P). A = 1 means 10% transmitted; A = 2 means 1%.

Beer–Lambert law. For a dilute solution of one absorbing species in monochromatic light, A = ε·l·c, where ε is the molar absorptivity (a property of the substance at that wavelength), l the path length and c the concentration. Absorbance, not transmittance, is linear in concentration, so calibration curves plot A against c. Absorbances of several non-interacting species at one wavelength add, which allows mixtures to be analysed by measuring at as many wavelengths as there are components.

Deviations from Beer's law. High concentrations (solute interactions, refractive-index change), chemical equilibria (dissociation, association), polychromatic light (finite bandwidth on a sloping absorption band), stray light reaching the detector, scattering by turbid samples and fluorescence all bend the calibration curve. Stray light is especially damaging at high absorbance, putting a ceiling on measurable A.

Instrument components.

  • Sources: deuterium lamp for UV (about 190–350 nm), tungsten-halogen lamp for visible and near IR (about 350–2500 nm), xenon flash lamps, LEDs in simple photometers.
  • Wavelength selector: absorption or interference filters (photometers/colorimeters), or a monochromator — entrance slit, collimator, dispersing element (prism or diffraction grating), focusing optics and exit slit. Narrower slits give finer spectral bandwidth but less light.
  • Sample cells (cuvettes): glass or plastic for the visible, fused silica (quartz) for the UV, standard path 1 cm.
  • Detectors: photomultiplier tubes (high gain, low light), silicon photodiodes, photodiode or CCD arrays (whole spectrum at once in diode-array instruments).
  • Readout: a ratio of sample to reference signals, converted to A.

Single-beam and double-beam. A single-beam instrument measures the blank (solvent) and then the sample, so source drift between the two readings appears as error. A double-beam instrument splits or chops the light between sample and reference cells and takes their ratio continuously, cancelling source and detector drift.

Diffraction grating. A reflection grating with groove spacing d sends wavelength λ in order m to the angle satisfying m·λ = d(sin θ_i + sin θ_m); at normal incidence m·λ = d·sin θ. Its resolving power is R = λ/Δλ = m·N, where N is the number of illuminated grooves.

Photometric accuracy. A fixed error in reading T gives the smallest relative concentration error near A ≈ 0.434 (T ≈ 37%); in practice keep A between about 0.2 and 1.0 by diluting or changing path length.

Other optical measurements. Optical power meters (calibrated photodiodes or thermal heads, readings in dBm), optical spectrum analysers (scanning grating monochromators), colorimeters, turbidimeters and nephelometers (scattered light), fluorometers and flame photometers apply the same source–selector–detector chain.

Formulas

T = P/P₀, A = −log₁₀T = log₁₀(P₀/P) P₀ incident and P transmitted radiant power (W) or intensity (W/m²); T and A dimensionless.

A = ε·l·c ε molar absorptivity (L·mol⁻¹·cm⁻¹), l path length (cm), c concentration (mol/L).

A_total = Σ ε_i·l·c_i Additivity for non-interacting species.

m·λ = d·sin θ (normal incidence) m diffraction order, d groove spacing (m), θ diffraction angle.

R = λ/Δλ = m·N R resolving power, Δλ smallest resolvable wavelength difference, N number of illuminated grooves.

T_measured = (T + s)/(1 + s) s stray-light fraction (relative to P₀).

Worked examples

Example 1 (standard). A dye with ε = 1.2 × 10⁴ L·mol⁻¹·cm⁻¹ at its absorption peak shows 40% transmittance in a 1 cm cell. Find the absorbance and concentration.

  1. A = −log₁₀(0.40) = 0.398.
  2. c = A/(ε·l) = 0.398/(1.2 × 10⁴ × 1).
  3. c = 3.32 × 10⁻⁵ mol/L. Answer: A ≈ 0.398, c ≈ 3.3 × 10⁻⁵ mol/L (33 µmol/L).

Example 2 (GATE level). (a) A grating with 600 lines/mm is illuminated at normal incidence over a width of 50 mm. Find the first-order angle for 589 nm and the smallest resolvable Δλ there. (b) A sample has true transmittance 1% (A = 2). The instrument has 1% stray light. What absorbance does it display?

  1. (a) d = 1/600 mm = 1.667 µm; sin θ = mλ/d = 0.589/1.667 = 0.3534, so θ = 20.7°.
  2. N = 600 × 50 = 30 000; Δλ = λ/(mN) = 589/30 000 = 0.0196 nm — easily resolving the 0.6 nm sodium doublet.
  3. (b) T_measured = (0.01 + 0.01)/(1 + 0.01) = 0.0198.
  4. A_measured = −log₁₀(0.0198) = 1.70. Answer: (a) 20.7°, 0.020 nm; (b) A reads 1.70 instead of 2.00 — a 15% negative error from only 1% stray light.

Example 3 (dilution). A sample reads A = 2.4, outside the reliable range. It is diluted 1:5 and re-measured. What absorbance is expected?

  1. A is proportional to c, so A_new = 2.4/5 = 0.48. Answer: about 0.48, within the 0.2–1.0 range; multiply the result by 5 afterwards.

Common mistakes

  • Treating transmittance as linear in concentration. Only absorbance is.
  • Using natural log instead of log₁₀ for absorbance.
  • Forgetting %T must be converted to a fraction (40% → 0.40).
  • Measuring UV absorbance in glass or plastic cuvettes, which themselves absorb below about 320 nm.
  • Ignoring the blank: solvent and cell absorbance must be subtracted (zeroed) first.
  • Trusting readings above A ≈ 2 when stray light and noise dominate.

For GATE IN

  • Transmittance, absorbance and Beer–Lambert concentration numericals, including mixtures at two wavelengths.
  • Grating equation and resolving power.
  • Effect of stray light and polychromatic light on measured absorbance.
  • Instrument blocks: sources and detectors for UV, visible and IR; single- versus double-beam.

Quick check

  1. What absorbance corresponds to T = 0.2?
  2. If path length doubles at constant concentration, what happens to A?
  3. Which source covers the UV region in a UV–visible spectrophotometer?
  4. A 1200 lines/mm grating at normal incidence, first order, 500 nm: what is θ? Answers: 1. 0.699. 2. It doubles. 3. Deuterium lamp. 4. 36.9°.

Try answering each one aloud before you open it.

  1. 1.What is spectrophotometry?Concept

    Spectrophotometry is a method used to measure how much a chemical substance absorbs light by measuring the intensity of light as a beam of light passes through a sample solution. It is commonly used in quantitative analysis in various fields such as chemistry, physics, biochemistry, and molecular biology.

  2. 2.Explain the basic principle of a spectrophotometer.Concept

    A spectrophotometer works on the principle of Beer-Lambert Law, which states that the absorbance of light by a substance is directly proportional to its concentration and the path length of the light through the sample. The device measures the intensity of light before and after it passes through a sample, and the difference in intensity is used to calculate the absorbance.

  3. 3.What are the main components of a spectrophotometer?Concept

    The main components of a spectrophotometer include a light source, a monochromator or filter to select the desired wavelength, a sample holder, a detector to measure the intensity of transmitted light, and a digital display or computer to show the results. Each component plays a crucial role in ensuring accurate measurements.

  4. 4.Why is a monochromator used in a spectrophotometer?Application

    A monochromator is used in a spectrophotometer to isolate a specific wavelength of light from a broad spectrum. This is important because different substances absorb light at different wavelengths, and isolating a specific wavelength allows for accurate measurement of the absorbance of the substance being analyzed.

  5. 5.What happens if the path length of the cuvette is doubled in a spectrophotometric measurement?Application

    If the path length of the cuvette is doubled, the absorbance of the sample will also double, assuming the concentration of the sample remains constant. This is because absorbance is directly proportional to the path length according to the Beer-Lambert Law.

  6. 6.How does temperature affect spectrophotometric measurements?Application

    Temperature can affect spectrophotometric measurements by altering the refractive index of the solvent, changing the solubility of the solute, or causing chemical reactions that change the concentration of the absorbing species. It is important to maintain a constant temperature during measurements to ensure accuracy.

  7. 7.What is the significance of the Beer-Lambert Law in spectrophotometry?Concept

    The Beer-Lambert Law is significant in spectrophotometry because it provides a mathematical relationship between absorbance, concentration, and path length. This law allows for the quantitative determination of the concentration of an unknown sample by measuring its absorbance at a specific wavelength.

  8. 8.Calculate the concentration of a solution if the absorbance is 0.5, the molar absorptivity is 100 L/mol·cm, and the path length is 1 cm.Numerical

    Using the Beer-Lambert Law, A = ε·c·l, where A is absorbance, ε is molar absorptivity, c is concentration, and l is path length. Rearranging for concentration, c = A / (ε·l). Substituting the given values, c = 0.5 / (100 L/mol·cm × 1 cm) = 0.005 mol/L.

  9. 9.What is the role of a detector in a spectrophotometer?Concept

    The detector in a spectrophotometer measures the intensity of light that passes through the sample. It converts the light signal into an electrical signal, which is then processed to determine the absorbance of the sample. Common types of detectors include photodiodes and photomultiplier tubes.

  10. 10.If a spectrophotometer shows a negative absorbance reading, what could be the possible reasons?Application

    A negative absorbance reading in a spectrophotometer could be due to instrument calibration errors, incorrect baseline setting, or stray light affecting the measurement. It may also occur if the sample is more transparent than the reference, indicating an issue with the sample or the setup.

Finished this topic? Mark it so your progress, study plan and readiness keep up.

Stuck on something here?