Electronic instruments: multimeter and oscilloscope

Measuring and analyzing electrical signals with multimeters and oscilloscopes, including principles, meter design (shunt/multiplier), loading effect, CRO/DSO operation, probes, bandwidth and rise time, and typical exam problems.

Drafted with Aria, reviewed by the AiCanCode.org team · Updated . How we write and check our notes. Spotted an error? Use Give Feedback at the bottom of the page.

Why it matters

Engineers measure voltage, current, resistance, and waveforms in labs, field testing, maintenance, and debugging circuits. Multimeters give accurate numeric values; oscilloscopes show how signals change with time, revealing faults like noise, distortion, and timing errors.

Key ideas

  • Digital Multimeter (DMM) — a multimeter with a digital display. Internally it includes an analog-to-digital converter (ADC), input attenuators, protection (fuses, MOVs), and firmware to compute RMS for AC. Advantages: high input impedance (typically 10 MΩ on volts), accuracy, no parallax, extra functions.

  • Analog multimeter (VOM) — a meter with a moving-coil meter movement and a needle. It uses either a permanent-magnet moving-coil (PMMC) movement (for DC) with rectifier (for AC). Key parameter: sensitivity in Ω/V (ohms per volt).

  • Sensitivity (of analog voltmeter) — exact definition: Sensitivity S is the input resistance per volt of full-scale reading; S = Rin/Vfs and is expressed in Ω/V. For a PMMC movement with full-scale current Ifs, S = 1/Ifs (Ω/V). Assumption: linear meter movement, negligible rectifier drop on DC range.

  • Multiplier resistor (voltmeter design) — a series resistor added to a meter movement so that a higher full-scale voltage can be measured without exceeding Ifs. Design equation derives from Ohm’s law: Rmult = Vrange/Ifs − Rm, where Rm is movement resistance.

  • Shunt resistor (ammeter design) — a low-value resistor placed in parallel with a meter movement so that most current bypasses the coil. Design equation: Rsh = (Ifs·Rm)/(Irange − Ifs). Assumption: ideal connections, temperature constant.

  • Loading effect — when a meter draws current from the circuit, it changes the quantity being measured, causing error. A voltmeter with finite input resistance forms a parallel path with the circuit resistance. Larger Rin reduces loading; DMMs (≈10 MΩ) load far less than analog meters (e.g., 20 kΩ/V on DC ranges).

  • Derivation for loading, typical case (what to draw): Draw a Thevenin source Vth with series resistance Rs feeding a node where the voltmeter Rin is connected to ground. Without meter, node voltage is Vth. With meter, voltage becomes V = Vth·(Rin/(Rs + Rin)). The percentage error = [(Vth − V)/Vth]×100 = [Rs/(Rs + Rin)]×100. Label Vth, Rs, Rin, V.

  • AC measurement and RMS — exact: Root-Mean-Square (RMS) value of a periodic voltage v(t) of period T is Vrms = sqrt((1/T)∫0→T v(t)^2 dt). Many DMMs report true RMS for any waveform within crest factor and bandwidth limits; some cheaper meters report the “RMS-equivalent” of the average-rectified value assuming a pure sine wave (gives correct RMS for sine only).

  • Oscilloscope — an instrument that displays voltage as a function of time on a screen. It has vertical (Y) system for input signal and scale (volts/div), horizontal (X) time base with sweep generator (time/div), trigger system for stable display, and a display (CRT in CRO, LCD in DSO).

  • Cathode Ray Oscilloscope (CRO) — analog scope using a Cathode Ray Tube. Key blocks: CRT (electron gun, deflection plates, phosphor screen), vertical amplifier, time-base generator, trigger, and power supply. What to draw: Block diagram with signal path from input to vertical plates; time-base to horizontal plates; trigger controlling sweep; CRT labeled gun, deflection plates, and screen; grounding shown.

  • Digital Storage Oscilloscope (DSO) — samples the input via an ADC at sampling frequency fs, stores samples in memory, and displays the waveform. Advantages: storage, measurement cursors, FFT, higher input impedance with probes, automatic measurements, screenshots. Limitation: aliasing if fs is too low.

  • Bandwidth (oscilloscope) — exact: The bandwidth BW is the frequency at which the instrument’s sinusoidal response falls to −3 dB (output is 0.707 of low-frequency amplitude) relative to low-frequency gain. For a single-pole system, step response rise time tr (10%–90%) approximately follows tr ≈ 0.35/BW. Assumptions: single-pole dominant response, bandwidth in Hz, tr in seconds.

  • Probe — an accessory that connects the circuit to the scope. A 1× probe presents ≈1 MΩ ∥ ~100 pF; a 10× attenuating probe presents ≈10 MΩ ∥ ~10–15 pF and scales the signal by 1/10. The 10× probe reduces loading and extends high-frequency performance if compensated.

  • Probe compensation — a small adjustable capacitor in the probe forms an RC divider matching the input capacitance of the scope so that attenuation is flat with frequency. What to draw: RC divider with Rprobe in series; Cprobe (adjustable) in parallel with Rprobe; and scope input Rin∥Cin to ground; show test-square-wave input and under/over-compensation waveforms (label undershoot/rounded, overshoot/peaked).

  • Triggering (scope) — system that starts each sweep at a defined condition (e.g., rising edge at set level) to produce a stable waveform. Modes: auto, normal, single, with sources CH1/CH2/Ext and couplings AC/DC.

  • Time base — sawtooth or ramp generator in CRO (or equivalent sampling-time control in DSO) that moves the spot horizontally at a constant rate. Setting is time per division (s/div). Assumes calibrated sweep.

  • Lissajous figures — with CRO X–Y mode (disable time base), apply one sinusoid to X and another to Y. The resulting ellipse allows measurement of phase difference and frequency ratio. For phase φ when fx = fy and amplitudes equal, sin|φ| = b/a where a and b are the axis intercepts (x = ±a, y = ±b); these are not the ellipse’s principal axes. What to draw: axes X and Y with ellipse touching ±a on X and ±b on Y.

  • Safety categories and fuses — DMM inputs have protection ratings (CAT II/III/IV) and internal fuses on current ranges. Always insert leads correctly and de-energize circuits when measuring resistance/continuity.

Formulas

  • Rsh = (Ifs·Rm)/(Irange − Ifs) — Shunt resistor for an ammeter. Irange is desired full-scale current (A). Applies when meter movement is in parallel with shunt.

  • V = Vth·(Rin/(Rs + Rin)) — Loaded voltage with a voltmeter. Vth: Thevenin source voltage (V); Rs: source resistance (Ω); Rin: meter input resistance (Ω). Applies to DC or AC at low frequency where resistive model holds.

  • Error% = [Rs/(Rs + Rin)] × 100 — Percentage loading error relative to the true open-circuit voltage Vth. Same assumptions as above.

  • Vrms = sqrt((1/T)·∫0→T v(t)^2 dt) — RMS definition; v(t) in volts, T in seconds. Applies to periodic signals.

  • tr ≈ 0.35/BW — Oscilloscope system rise time–bandwidth relation. tr in seconds; BW in Hz. Single-pole approximation.

  • tr_total = sqrt(tr1² + tr2²) — Combined rise time (10–90%) for cascaded single-pole/Gaussian limits.

  • tr_signal ≈ sqrt(tr_meas² − tr_instr²) and BW_signal ≈ 0.35/tr_signal — De-embed instrument rise time, then convert to bandwidth using the single-pole approximation.

  • fs ≥ 2·fmax — Sampling theorem condition to avoid aliasing (Nyquist). fs sampling rate (samples/s); fmax highest frequency content (Hz). Practical DSO recommendation: fs ≥ 5–10× highest frequency to capture shape.

  • Phase via Lissajous (equal freq): sin|φ| = b/a — a is the X-axis intercept magnitude; b is the Y-axis intercept magnitude (equal amplitudes, same frequency). Direction of rotation indicates lead/lag.

  • f = 1 / (Ndiv × time_per_div) — Ndiv: horizontal divisions per cycle (dimensionless); time_per_div in s/div; result in Hz.

Worked examples

  1. Designing an analog voltmeter range (multiplier resistor) Given: A PMMC movement has Ifs = 50 µA and Rm = 2 kΩ. Design a 0–10 V DC range. Steps:
  1. Formula: Rmult = Vrange/Ifs − Rm.
  2. Numbers: Vrange = 10 V; Ifs = 50 µA = 50 × 10⁻⁶ A; Rm = 2000 Ω.
  3. Compute Vrange/Ifs = 10 / (50 × 10⁻⁶) = 200000 Ω = 200 kΩ.
  4. Subtract Rm: Rmult = 200 kΩ − 2 kΩ = 198 kΩ. Answer: The series multiplier resistor is approximately 198 kΩ. Choose nearest standard value 200 kΩ for slight under-range (verify tolerance in design context). Final: Rmult = 198 kΩ (ideal).
  1. Designing an analog ammeter range (shunt resistor) Given: Same PMMC movement (Ifs = 50 µA, Rm = 2 kΩ). Design a 0–10 mA DC ammeter. Steps:
  1. Formula: Rsh = (Ifs·Rm)/(Irange − Ifs).
  2. Numbers: Ifs = 50 × 10⁻⁶ A; Rm = 2000 Ω; Irange = 10 × 10⁻³ A.
  3. Compute numerator: Ifs·Rm = 50 × 10⁻⁶ × 2000 = 0.1 V (equivalently 0.1 Ω·A).
  4. Denominator: Irange − Ifs = 0.01 − 0.00005 = 0.00995 A.
  5. Rsh = 0.1 / 0.00995 ≈ 10.05 Ω. Answer: Use a shunt of approximately 10.05 Ω. In practice, choose a low-TC resistor near 10 Ω and calibrate.
  1. Loading effect of analog vs digital voltmeter Given: Measure the open-circuit voltage Vth = 12.0 V across a source with Rs = 50 kΩ. Compare readings with:
  • Analog voltmeter sensitivity S = 20 kΩ/V on 10 V range (Rin = S × Vrange = 200 kΩ). We will still attempt 10 V range knowing 12 V exceeds it, to illustrate; then try 15 V range if available.
  • DMM with Rin = 10 MΩ on any DC volts range. Steps: A) For analog 10 V range (incorrect range for 12 V; over-range): The pointer would hit the end stop; reading is unreliable. Correct approach is to choose a higher range. So select 15 V range (common on analog meters). For 15 V range: Rin = 20 kΩ/V × 15 V = 300 kΩ. Formula: V = Vth·(Rin/(Rs + Rin)). Numbers: V = 12 × (300k / (50k + 300k)) = 12 × (300/350) = 12 × 0.8571 = 10.285 V. Error% = [(12 − 10.285)/12] × 100 ≈ 14.29%. B) For DMM: Rin = 10 MΩ. V = 12 × (10M / (50k + 10M)) ≈ 12 × (10,000k / 10,050k) ≈ 12 × 0.9950 = 11.94 V. Error% ≈ 0.50%. Answer: Analog meter reading ≈ 10.29 V (≈14.3% low); DMM reading ≈ 11.94 V (≈0.5% low).
  1. Reading amplitude, period, and frequency on a DSO; estimating rise time Given: A repetitive square wave is displayed. Vertical scale is 2 V/div; horizontal scale is 0.5 ms/div. The waveform spans 3 vertical divisions peak-to-peak and 4 horizontal divisions per cycle. The scope bandwidth is 20 MHz. Steps:
  1. Peak-to-peak voltage: Vpp = 3 div × 2 V/div = 6 V. For a symmetric square, Vp = Vpp/2 = 3 V.
  2. Period: T = 4 div × 0.5 ms/div = 2.0 ms.
  3. Frequency: f = 1/T = 1 / 2.0 ms = 500 Hz.
  4. Bandwidth–rise time estimate: tr ≈ 0.35/BW = 0.35 / 20×10⁶ = 17.5 ns. This is the fastest edge the 20 MHz scope can display (instrument limit). The measured rise time of the signal will be tr_meas ≈ sqrt(tr_signal² + tr_scope²). If you measure tr_meas =, say, 50 ns, then tr_signal ≈ sqrt(50² − 17.5²) ns ≈ 46.8 ns. Answer: Vpp = 6 V, f = 500 Hz, tr(scope limit) ≈ 17.5 ns; example de-embedded signal rise time ≈ 46.8 ns if 50 ns observed.
  1. Phase by Lissajous (equal frequencies) Given: In X–Y mode with equal amplitude sinusoids from same-frequency sources, the ellipse touches ±4.0 V on X (a = 4.0 V) and ±2.0 V on Y (b = 2.0 V). Find phase difference. Steps:
  1. Formula: sin|φ| = b/a (for equal frequencies and equal amplitudes; here the equal amplitude condition is met on-screen by equal X and Y scaling so that intercepts represent equal input peaks). Here a = 4.0, b = 2.0.
  2. Compute: sin|φ| = 2.0/4.0 = 0.5 ⇒ |φ| = 30°.
  3. Sign (lead/lag) is determined by observing rotation with time: if the trace moves clockwise with time, Y lags X; counter-clockwise indicates Y leads. Assume clockwise ⇒ Y lags by 30°. Answer: Phase difference magnitude 30° (Y lags X if clockwise motion observed).

Common mistakes

  • Ignoring loading: using an analog meter with 20 kΩ/V on high-resistance circuits, causing large measurement error.
  • Confusing RMS, average, and peak values; using Vpp/2√2 for a square wave (wrong; square-wave Vrms = Vp, not Vp/√2).
  • Misapplying tr ≈ 0.35/BW to non-single-pole systems or to digital signals with significant bandwidth limiting in the source; not combining rise times in quadrature.
  • Poor probe compensation leading to rounded (under-compensated) or overshoot (over-compensated) square waves and wrong amplitude readings.
  • Grounding errors with CRO/DSO earth-referenced ground clip causing short-circuits in mains-referenced circuits. Use differential probes or isolation where appropriate.

For your exam

  • Derivations: voltmeter multiplier resistor; ammeter shunt resistor; loading error using Thevenin equivalent; rise time–bandwidth relation for a single-pole system (start with first-order step response v(t) = V(1 − e^(−t/τ)), define tr from 10% to 90%, derive tr ≈ 2.2τ and BW = 1/(2πτ), thus tr ≈ 0.35/BW).
  • Numericals: compute shunt/multiplier values; compute loading error; compute input resistance from sensitivity and range; read oscilloscope amplitude/frequency; use rise time and bandwidth to estimate limits; Lissajous phase.
  • Full-marks answers include: correct formula stated first; clearly labeled diagram with all parameters (units, ranges, divisions); assumptions (e.g., single-pole response, ideal meter movement); stepwise calculation with units; final numeric answer highlighted; brief discussion of sources of error and instrument limits.

Quick check

  1. A PMMC movement has Ifs = 100 µA and Rm = 1 kΩ. What multiplier is needed for a 0–5 V voltmeter range?
  2. A DMM with 10 MΩ input measures a source Vth = 5.00 V with Rs = 100 kΩ. What is the indicated voltage and percentage error?
  3. A 100 MHz oscilloscope has what approximate 10–90% rise time limit? Answers: 1) Rmult = 5/(100 µA) − 1 kΩ = 50 kΩ − 1 kΩ = 49 kΩ. 2) V = 5 × (10M/(100k + 10M)) ≈ 5 × 0.9901 = 4.950 V; Error ≈ 0.99%. 3) tr ≈ 0.35/100 MHz = 3.5 ns.