Composition analysers and transmitters
On-line composition analysers (GC, IR, oxygen, pH), sample conditioning and analyser dead time, and how transmitters convert measurements into standard signals.
Drafted with Aria, reviewed by the AiCanCode.org team. Spotted an error? Use Give Feedback at the bottom of the page.
Why it matters
Product quality is defined by composition, not by temperature or pressure, so the most valuable loops in a plant (distillation purity, reactor conversion, flue-gas oxygen, effluent pH) end at an analyser. Analysers are also the slowest, least reliable and most expensive instruments on the plant, and their delays shape how tightly a composition loop can be tuned.
Key ideas
Direct and inferential measurement. A composition can be measured directly by an on-line analyser, or inferred from easier measurements (for example a tray temperature in a distillation column at known pressure, or density/refractive index of a binary mixture). Inferential measurements are fast but drift when the third components or the pressure change.
Common on-line analysers.
- Gas chromatograph (GC): a sample is injected into a carrier gas and separated in a column by differences in partitioning with the stationary phase; components elute at characteristic retention times and are detected by a thermal-conductivity (TCD) or flame-ionisation (FID) detector. Peak area is proportional to amount after calibration. It is a sampled (discontinuous) analyser with a cycle time of minutes.
- Infrared (IR/NDIR) and UV analysers: molecules absorb at characteristic wavelengths; the absorbance follows the Beer–Lambert law. NDIR is standard for CO, CO₂ and hydrocarbons.
- Oxygen analysers: paramagnetic (O₂ is strongly attracted by a magnetic field) and zirconia cells (a solid electrolyte at high temperature produces an emf depending on the ratio of O₂ partial pressures, Nernst equation), used for combustion control.
- Thermal-conductivity analysers for binary gases such as H₂ in N₂.
- Electrochemical: pH (glass electrode, Nernst slope about 59.16 mV per pH unit at 25 °C), ORP, conductivity, dissolved oxygen.
- Physical-property analysers: density, refractive index, viscosity, boiling point and flash point for products.
- Mass spectrometer: ionises the sample and sorts ions by mass-to-charge ratio; fast and multi-component but costly.
Sample conditioning. Most analysers need a sample withdrawn through a probe and a transport line, then filtered, pressure-reduced, cooled or vaporised, and returned or vented. The transport line adds a pure dead time equal to line volume divided by sample flow, often the largest delay in the loop. Fast-loop (bypass) sampling keeps this short.
Analyser dead time. A sampled analyser with cycle time T reports a result at the end of each cycle and holds it until the next. The information is therefore between T and 2T old, so the analyser behaves roughly like a dead time of about 1.5 T plus the sample transport delay. Dead time limits achievable controller gain (see controller tuning).
Transmitters. A transmitter converts the primary element's signal into a standard signal: 4–20 mA, 20–100 kPa pneumatic, or digital (HART superimposed on 4–20 mA, Foundation Fieldbus, Profibus). Key settings are zero (the value at 4 mA) and span (the range covered). Smart transmitters allow remote re-ranging, diagnostics and temperature compensation. Elevated or suppressed zero is used, for example, in level transmitters with wet legs.
Calibration. Analysers drift, so they are checked against certified calibration gases or standard solutions, usually at zero and one span point. A two-point calibration fixes both the offset and the slope of a linear response.
Formulas
A = −log₁₀(I/I₀) = ε·b·c
- Beer–Lambert law: A absorbance (dimensionless), I/I₀ transmittance, ε molar absorptivity (m²/mol), b path length (m), c concentration (mol/m³). Valid for dilute samples and monochromatic light.
c = c_std·(A / A_std)
- One-point calibration in the linear (Beer–Lambert) range with the same cell.
x = LRV + (I − 4)·(URV − LRV)/16
- Value x from a 4–20 mA signal I (mA) for a linear transmitter.
θ_line = V_line / Q_sample = (π·d²/4)·L / Q_sample
- Transport dead time (s): d tube inside diameter (m), L length (m), Q_sample sample flow (m³/s).
θ_analyser ≈ θ_line + 1.5·T_cycle
- Approximate effective dead time of a sampled analyser with cycle time T_cycle (s).
E = s·(7 − pH), s = 2.303·R·T/F ≈ 0.1984·T mV per pH
- Ideal glass electrode with isopotential point at pH 7: E in mV, T in K, R = 8.314 J/(mol·K), F = 96 485 C/mol. s = 59.16 mV/pH at 298.15 K.
Worked examples
Example 1 (standard): IR absorbance. An IR analyser cell gives an absorbance of 0.30 with a standard containing 2.0 mol/m³ of CO₂. A process sample in the same cell transmits 40 % of the incident beam. Find the CO₂ concentration.
- A = −log₁₀(I/I₀) = −log₁₀(0.40) = 0.398.
- Beer–Lambert, same ε and b: c = c_std·(A/A_std) = 2.0·(0.398/0.30).
- c = 2.65 mol/m³.
c ≈ 2.65 mol/m³
Example 2 (GATE level): dead time of a GC composition loop. A GC analysing a distillate has a cycle time of 4 min. The sample travels through 30 m of tubing of 6 mm inside diameter at 1.0 L/min. Estimate the effective dead time contributed by the measurement.
- Line volume: V = (π·d²/4)·L = (π·0.006²/4)·30 = 8.48 × 10⁻⁴ m³.
- Sample flow: Q = 1.0 L/min = 1.0 × 10⁻³/60 = 1.667 × 10⁻⁵ m³/s.
- Transport delay: θ_line = V/Q = 8.48 × 10⁻⁴ / 1.667 × 10⁻⁵ = 50.9 s.
- Sampling delay: 1.5·T_cycle = 1.5·240 = 360 s.
- θ ≈ 50.9 + 360 = 410.9 s.
θ ≈ 411 s (about 6.8 min). The analyser, not the column, may set the loop speed; a tray-temperature inner loop or a faster analyser is the usual remedy.
Example 3 (short): pH at temperature. At 60 °C (333.15 K) the electrode slope is s = 0.1984·333.15 = 66.1 mV/pH. A reading of E = +120 mV gives pH = 7 − 120/66.1 = 5.18. Using the 25 °C slope would give 7 − 120/59.16 = 4.97, which is why pH transmitters carry automatic temperature compensation.
Common mistakes
- Treating a GC like a continuous instrument and ignoring its cycle time and sample-line delay in the loop.
- Using percentage transmittance directly in place of absorbance; Beer–Lambert is linear in A = −log₁₀(I/I₀), not in I/I₀.
- Forgetting that a tray temperature is a composition proxy only at constant pressure.
- Mixing up zero and span when re-ranging a transmitter; changing the zero shifts the whole range, changing the span changes the slope.
- Applying the 25 °C pH slope at other temperatures.
For GATE CH
This topic mostly appears as conceptual questions: which analyser for which species (paramagnetic for O₂, NDIR for CO₂, GC for multicomponent hydrocarbons), what a transmitter does, and the 4–20 mA conversion. Its main numerical link is dead time: transport and sampling delays enter transfer functions as e^(−θs) and reduce the stability margin, which GATE tests through Bode and tuning questions.
Quick check
- Why is a gas chromatograph called a discontinuous analyser?
- A 4–20 mA O₂ transmitter ranged 0–5 % reads 14.4 mA. What is the O₂ content?
- Doubling the sample flow in a transport line changes the transport delay by what factor?
- Which property of oxygen does a paramagnetic analyser use?
Answers: 1. It analyses one injected sample per cycle and holds the result until the next cycle. 2. 3.25 % O₂. 3. It halves the delay. 4. Its strong attraction into a magnetic field (paramagnetism).
Interview questions
All Process Instrumentation and Control interview questionsTry answering each one aloud before you open it.
1.What is a composition analyser in the context of process instrumentation?Concept
A composition analyser is a device used in process industries to determine the chemical composition of a substance. It provides real-time data on the concentration of various components within a mixture, which is crucial for process control and optimization.
2.Explain the working principle of a gas chromatograph used as a composition analyser.Concept
A small sample is injected into a carrier gas and swept through a column; components partition differently between the carrier and the stationary phase, so they elute at different retention times. A detector such as a thermal-conductivity (TCD) or flame-ionisation (FID) detector produces a peak for each component, and the peak area, after calibration with a standard, gives its concentration. A process GC is a sampled analyser: it produces one result per cycle of several minutes, which adds significant dead time to a composition control loop.
3.What is a transmitter in process control, and how does it relate to composition analysers?Concept
A transmitter converts the primary sensor or analyser signal into a standard signal, usually 4–20 mA (often with HART digital data superimposed) or a fieldbus message, that the controller or DCS can read over long distances. It is set by its zero (value at 4 mA) and span (range covered); the live zero of 4 mA lets a broken loop be detected. An analyser normally has its own transmitter output so its composition reading can be used for indication, alarms and control.
4.Why are composition analysers critical in the petrochemical industry?Application
Composition analysers are critical in the petrochemical industry because they provide real-time data on the chemical composition of products and intermediates. This information is essential for ensuring product quality, optimizing processes, reducing waste, and maintaining safety standards.
5.What happens if a composition analyser in a chemical plant fails to provide accurate data?Application
If a composition analyser fails to provide accurate data, it can lead to incorrect process adjustments, resulting in off-spec products, increased waste, potential safety hazards, and financial losses. It may also affect compliance with regulatory standards.
6.How does a non-dispersive infrared (NDIR) analyser work for gas composition analysis?Concept
An NDIR analyser passes broadband infrared light through a sample cell and uses an optical filter (or a gas-filled detector) tuned to the absorption band of the target gas, such as CO₂ or CO, without dispersing the light into a spectrum. The reduction in transmitted intensity gives the absorbance, which by the Beer–Lambert law is proportional to concentration and path length. A reference cell or reference wavelength compensates for source drift and dirty windows.
7.Why is it important to calibrate composition analysers regularly?Application
Regular calibration of composition analysers is important to ensure accuracy and reliability of the data they provide. Calibration adjusts the analyser to account for any drift or changes in sensitivity over time, which can be caused by environmental factors or wear and tear.
8.Calculate the concentration of a component in a mixture if the peak area in a gas chromatogram is 1500 units and the response (calibration) factor is 0.5 area units per ppm.Numerical
For a linear detector response, concentration = peak area / response factor = 1500 / 0.5 = 3000 ppm. The response factor itself comes from injecting a standard of known concentration, which is why the calibration must be repeated when the detector drifts.
9.Explain how a mass spectrometer can be used as a composition analyser.Concept
A mass spectrometer ionizes chemical species and sorts the ions based on their mass-to-charge ratio. By analyzing the resulting mass spectrum, the composition of the sample can be determined, identifying and quantifying the different components present.
Finished this topic? Mark it so your progress, study plan and readiness keep up.
Stuck on something here?