Pressure Measurement

  • p_abs = p_gauge + p_atm — absolute pressure from gauge reading; all in Pa.
  • Δp = ρ·g·h — manometer with gas above both legs; ρ = manometer liquid density (kg/m³), g = 9.81 m/s², h = vertical column difference (m).
  • Δp = (ρm − ρf)·g·h — U-tube with process liquid (density ρf) above the manometer liquid (ρm).
  • Δp = ρ·g·L·(sinθ + a/A) — inclined well-type manometer; L = length moved along the tube (m), θ = tube angle from horizontal, a/A = tube-to-well area ratio.
  • x = p·Ae / k — bellows or spring-opposed element; x = deflection (m), Ae = effective area (m²), k = spring rate (N/m).
  • y₀ = 3·p·R⁴·(1 − ν²) / (16·E·t³) — centre deflection of a clamped flat diaphragm; R = radius, t = thickness (m), E = Young's modulus (Pa), ν = Poisson's ratio; valid for y₀ below about 0.3·t.

Vacuum Measurement

  • p ≈ ρ·g·a·h² / V — McLeod gauge, square-law method; a = sealed-capillary bore area (m²), h = compressed column length (m), V = trapped volume (m³), ρ = mercury density.
  • Ii = S·Ie·p — hot-cathode ionisation gauge; Ii = ion current (A), Ie = electron emission current (A), S = gas-dependent sensitivity (Pa⁻¹).

Differential Pressure Transmitters

  • I = 4 + 16·(x − LRV)/(URV − LRV) — 4–20 mA output for input x between the lower and upper range values.
  • Q = Q_max·√((I − 4)/16) — flow from a linear-DP transmitter on a head flow meter (LRV = 0).
  • ΔP = ρ·g·h − ρs·g·H — closed-tank level with a wet leg of seal liquid ρs and height H on the low side (zero elevation).

Head Flow Meters

  • Q = A·v — volumetric flow (m³/s) from area (m²) and mean velocity (m/s).
  • Q = Cd·A₂·√(2·ΔP/ρ) / √(1 − β⁴) — orifice, nozzle or venturi; Cd = discharge coefficient (≈ 0.6 orifice, ≈ 0.98 venturi), A₂ = bore or throat area (m²), β = d/D, ΔP = tap differential (Pa), ρ = fluid density (kg/m³).
  • v = √(2·(p₀ − p)/ρ) — Pitot tube point velocity; p₀ = stagnation pressure, p = static pressure (Pa).

Variable Area Flow Meters and Rotameters

  • ΔP = Vf·(ρf − ρ)·g / Af — constant pressure drop across the float; Vf = float volume (m³), ρf = float density, ρ = fluid density (kg/m³), Af = float area (m²).
  • Q = Cd·Aa·√(2·ΔP/ρ) — flow through the annulus; Aa = (π/4)·(Dt² − Df²) (m²).
  • Q₂/Q₁ = √[(ρf − ρ₂)·ρ₁ / ((ρf − ρ₁)·ρ₂)] — volumetric correction for a fluid of different density at the same float position.

Electromagnetic Flow Meters

  • E = B·D·v — electrode voltage (V); B = flux density (T), D = pipe inner diameter (m), v = mean velocity (m/s).
  • Q = π·D·E / (4·B) — volumetric flow (m³/s).

Ultrasonic Flow Meters

  • v = L·(1/t_d − 1/t_u) / (2·cosθ) — transit-time velocity, independent of sound speed; L = path length (m), t_d, t_u = downstream and upstream transit times (s), θ = path angle to the pipe axis.
  • v ≈ c²·Δt / (2·L·cosθ) — approximate form for v ≪ c; Δt = t_u − t_d (s), c = speed of sound (m/s).
  • Δf = 2·f₀·v·cosθ / c — Doppler meter frequency shift (Hz).

Vortex Flow Meters

  • f = St·v / d — shedding frequency (Hz); St = Strouhal number (≈ 0.2), v = velocity (m/s), d = bluff-body width (m).

Turbine, PD and Coriolis Flow Meters

  • f = K·Q — turbine meter; K = K-factor (pulses/m³), f = pulse frequency (Hz), Q = flow (m³/s).
  • Q = Vc·n — positive displacement meter; Vc = volume per cycle (m³), n = cycles per second.
  • F_c = 2·Δm·ω·v — Coriolis force on a fluid element of mass Δm (kg) moving at v (m/s) in a frame rotating at ω (rad/s).
  • ṁ = FCF·Δt — Coriolis mass flow (kg/s) from the pick-off time lag Δt and the flow calibration factor.
  • ρ = (m_t/V)·[(f₀/f)² − 1] — density from tube resonant frequency; f₀ = empty-tube frequency, m_t = tube mass, V = tube volume.

Level Measurement

  • h = ΔP / (ρ·g) — hydrostatic level (m) from bottom differential pressure (Pa).
  • W_app = W − ρ·g·A·h — displacer apparent weight (N); A = displacer area (m²), h = immersed length (m).
  • ΔP = g·(ρ₁·h₁ + ρ₂·h₂) — two liquid layers between taps (interface measurement).

Capacitive, Ultrasonic, Radar and Radiation Level Gauges

  • C = 2π·ε₀·[εr·h + (L − h)] / ln(b/a) — coaxial capacitance probe; ε₀ = 8.854 × 10⁻¹² F/m, εr = liquid relative permittivity, L = probe length, h = level, a = rod radius, b = tube radius (m).
  • d = c·t / 2 — echo distance (m) for ultrasonic or pulse radar; t = round-trip time (s); level = mounting height − d.
  • c_air ≈ 331.3·√(1 + T/273.15) — speed of sound in air (m/s), T in °C.
  • I = I₀·e^(−μ·x) — gamma attenuation; μ = linear attenuation coefficient (m⁻¹), x = path in material (m).

Industrial Temperature Measurement

  • E(T, 0) = E(T, T_ref) + E(T_ref, 0) — thermocouple law of intermediate temperatures; use table emf values, not temperatures.
  • R_T = R₀·(1 + α·T) — Pt100 RTD (linear approximation); R₀ = 100 Ω, α = 0.00385 /°C, T in °C.
  • lag = τ·R — first-order sensor following a ramp of R (°C/s); τ = time constant (s).
  • f_s = St·V / d — thermowell wake frequency (Hz); St ≈ 0.22, V = fluid velocity (m/s), d = tip diameter (m); keep well below the natural frequency as the code requires.

Density Measurement

  • ρ = m / V — density (kg/m³) from mass (kg) and volume (m³).
  • ρ = ΔP / (g·H) — hydrostatic density from two submerged taps H (m) apart.

Viscosity Measurement

  • τ_s = μ·(du/dy) — Newton's law of viscosity; τ_s = shear stress (Pa), μ = dynamic viscosity (Pa·s), du/dy = shear rate (s⁻¹).
  • μ = F·y / (A·u) — parallel-plate form; F = force (N), y = gap (m), A = plate area (m²), u = plate velocity (m/s).
  • μ = 2·r²·(ρ_s − ρ)·g / (9·v) — falling-ball (Stokes) viscometer; r = ball radius (m), v = terminal velocity (m/s); valid for Re < about 0.1.
  • ν = μ / ρ — kinematic viscosity (m²/s).

Humidity Measurement

  • RH = (p_v / p_s)·100 — relative humidity (%); p_v = partial pressure of water vapour, p_s = saturation pressure at the dry-bulb temperature (Pa).
  • ω = 0.622·p_v / (p − p_v) — humidity ratio (kg water per kg dry air); p = total pressure (Pa).

pH Measurement

  • pH = −log₁₀ a(H⁺) — a(H⁺) = hydrogen-ion activity (≈ concentration in mol/L for dilute solutions).
  • S = 2.303·R·T / F — Nernst slope (V per pH); 59.16 mV/pH at 25 °C, proportional to absolute temperature T (K).

Conductivity Measurement

  • κ = G·K = K / R — conductivity (S/m); G = conductance (S), R = resistance (Ω), K = L/A = cell constant (m⁻¹).
  • κ₂₅ = κ_T / (1 + α·(T − 25)) — linear temperature compensation; α ≈ 0.02 /°C for many salt solutions.

Gas Chromatography

  • k = (t_R − t_M) / t_M — retention factor; t_R = retention time, t_M = dead time of an unretained gas.
  • N = 16·(t_R / W_b)² — theoretical plates; W_b = baseline peak width (same time unit).
  • R_s = 2·(t_R2 − t_R1) / (W_b1 + W_b2) — resolution; R_s ≥ 1.5 gives baseline separation.

Spectroscopic Instruments

  • A = log₁₀(I₀/I) = ε·c·l — Beer–Lambert law; ε = molar absorptivity (L/(mol·cm)), c = concentration (mol/L), l = path length (cm).
  • E = h·c / λ — photon energy (J); h = 6.626 × 10⁻³⁴ J·s, c = 2.998 × 10⁸ m/s, λ = wavelength (m).
  • n·λ = d·(sin i + sin θ) — diffraction grating; d = groove spacing (m), n = order.

Hazardous Areas and Calibration

  • P_o = U_o·I_o / 4 — maximum power from a resistive-limited IS barrier (W).
  • C_i + C_cable ≤ C_o, L_i + L_cable ≤ L_o — intrinsic-safety entity check.
  • p = m·g / A — deadweight tester pressure (Pa); m = total mass (kg), g = local gravity (m/s²), A = piston area (m²).
  • % error of span = (indicated − true) / span × 100 — calibration error.