Electrical and Electronic Measurements — formula sheet

Instrumentation Engineering · GATE IN · SI units. Drafted with Aria, checked by the AiCanCode.org team.

Measurement errors and statistical analysis

  • δA = Am − At — Static error; Am = measured value, At = true value (same unit). Static correction = −δA.
  • εr = δA / At — Relative error (×100 for %).
  • δA = ±εFS × full-scale value — Limiting error of an instrument guaranteed as a fraction εFS of full scale; % error at a reading Am = δA/Am × 100.
  • X = x₁ ± x₂ → δX = ±(δx₁ + δx₂) — Limiting error of a sum or difference: absolute errors add.
  • X = x₁ⁿ·x₂ᵐ → δX/X = ±(|n|·δx₁/x₁ + |m|·δx₂/x₂) — Limiting error of a product, quotient or power: relative errors add, multiplied by the exponents.
  • δX/X = √((δx₁/x₁)² + (δx₂/x₂)²) — Root-sum-square estimate for independent random errors of a product or quotient.
  • x̄ = Σxᵢ / n — Arithmetic mean of n readings.
  • D = Σ|xᵢ − x̄| / n — Average deviation.
  • σ = √(Σ(xᵢ − x̄)² / (n − 1)) — Standard deviation of a small sample (use n instead of n − 1 for a large population); variance V = σ².
  • r = 0.6745·σ — Probable error of one reading (Gaussian).
  • σm = σ / √n ; rm = r / √n — Standard deviation and probable error of the mean.

PMMC, moving iron and electrodynamometer instruments

  • Td = N·B·A·I — PMMC deflecting torque (N·m); N = turns, B = gap flux density (T), A = coil area (m²), I = current (A).
  • Tc = K·θ — Spring controlling torque; K = spring constant (N·m/rad or N·m/degree).
  • θ = N·B·A·I / K — PMMC deflection: linear scale, responds to average current.
  • Td = ½·I²·dL/dθ — Moving-iron torque; L = coil inductance (H), dL/dθ in H/rad; responds to RMS.
  • θ = (I² / 2K)·dL/dθ — Moving-iron deflection with spring control.
  • Td = I₁·I₂·cos φ·dM/dθ — Electrodynamometer mean torque; I₁, I₂ = RMS coil currents (A), φ = phase angle between them, M = mutual inductance (H).

Extension of ammeter and voltmeter ranges

  • m = I / Im ; Rsh = Rm / (m − 1) = Im·Rm / (I − Im) — Ammeter shunt; Im = meter full-scale current (A), Rm = meter resistance (Ω), I = required range (A).
  • m = V / (Im·Rm) ; Rs = (m − 1)·Rm = V / Im − Rm — Voltmeter multiplier (Ω); V = required range (V).
  • S = 1 / Im — Voltmeter sensitivity (Ω/V); total voltmeter resistance on a range = S × range.
  • Rx = Im·(Rm + Rt) / I — Ayrton shunt tap for range I; Rt = total shunt resistance = Rm/(m₁ − 1) for the lowest range.
  • V_read = Vth·Rv / (Rv + Rth) — Voltmeter loading; Vth, Rth = Thevenin equivalent seen by the meter, Rv = voltmeter resistance (Ω).

Measurement of low, medium and high resistance

  • Rm = R·Rv / (R + Rv) — Ammeter–voltmeter, voltmeter directly across R: reads low; suits low R.
  • Rm = R + Ra — Ammeter–voltmeter, voltmeter across ammeter and R: reads high; suits high R.
  • R = √(Ra·Rv) — Crossover resistance between the two connections; Ra, Rv = ammeter and voltmeter resistances (Ω).
  • R = t / (C·ln(V / v)) = 0.4343·t / (C·log₁₀(V / v)) — Loss-of-charge method; t = time (s), C = capacitance (F), V = initial and v = final voltage (V).

DC bridges: Wheatstone and Kelvin double bridge

  • R = S·P / Q — Wheatstone balance; P, Q = ratio arms, S = standard arm, R = unknown (Ω), with P, Q forming one leg and R, S the other.
  • Vth = E·[R / (R + S) − P / (P + Q)] — Open-circuit galvanometer voltage of an unbalanced bridge; E = supply (V).
  • Rth = P·Q / (P + Q) + R·S / (R + S) ; Ig = Vth / (Rth + Rg) — Thevenin resistance and galvanometer current; Rg = galvanometer resistance (Ω).
  • R = (P / Q)·S + [q·r / (p + q + r)]·(P / Q − p / q) — Kelvin double bridge; P, Q outer and p, q inner ratio arms, S = standard, r = link resistance (Ω).
  • R = (P / Q)·S — Kelvin double bridge with P/Q = p/q (link resistance eliminated).

AC bridges: Maxwell, Hay, Schering and Wien

  • Z1·Z4 = Z2·Z3 ; θ1 + θ4 = θ2 + θ3 — General AC balance (arms 1 and 4 opposite): magnitude and angle conditions.
  • L1 = R2·R3·C4 ; R1 = R2·R3 / R4 ; Q = ω·C4·R4 — Maxwell inductance–capacitance bridge (C4 ∥ R4 opposite the coil); suits 1 < Q < 10.
  • L1 = R2·R3·C4 / (1 + ω²C4²R4²) ; R1 = ω²C4²R2R3R4 / (1 + ω²C4²R4²) ; Q = 1 / (ω·C4·R4) — Hay bridge (C4 in series with R4); suits Q > 10.
  • C1 = C2·R4 / R3 ; r1 = R3·C4 / C2 ; tan δ = ω·C4·R4 — Schering bridge; C1, r1 = specimen series capacitance (F) and loss resistance (Ω), C2 = standard capacitor, R4 ∥ C4 = arm opposite the specimen.
  • f = 1 / (2π·√(R1·R2·C1·C2)) ; R3 / R4 = R1 / R2 + C2 / C1 — Wien bridge; R1–C1 series arm, R2 ∥ C2 parallel arm. With equal R and C: f = 1/(2πRC), R3 = 2R4.

Measurement of power and wattmeters

  • P = V·I·cos φ ; Q = V·I·sin φ ; S = V·I — Active (W), reactive (var) and apparent (VA) power; V, I = RMS values, φ = phase angle.
  • True power = reading × cos φ / (cos β·cos(φ − β)) ; β = tan⁻¹(ωL / Rp) — Correction for pressure-coil inductance (lagging load); L = PC inductance (H), Rp = PC circuit resistance (Ω).
  • P = √3·VL·IL·cos φ — Balanced three-phase power; VL = line voltage (V), IL = line current (A).
  • W1 = VL·IL·cos(30° − φ) ; W2 = VL·IL·cos(30° + φ) — Two-wattmeter readings for a balanced load.
  • P = W1 + W2 ; Q = √3·(W1 − W2) ; tan φ = √3·(W1 − W2) / (W1 + W2) — Two-wattmeter method (P = W1 + W2 holds for any 3-wire load).
  • Q = √3 × W — Reactive power of a balanced load from one wattmeter with CC in one line and PC across the other two.

Measurement of energy and energy meters

  • E = P·t ; 1 kWh = 3.6 × 10⁶ J — Energy; P in kW and t in h give kWh.
  • E_registered = N / K — N = revolutions or pulses, K = meter constant (rev/kWh or imp/kWh).
  • % error = (E_registered − E_true) / E_true × 100 — Energy-meter error from a timed test.
  • Td ∝ V·I·sin(Δ − φ) — Induction-meter driving torque; Δ = angle by which the shunt flux lags V (ideally 90°), φ = load phase angle.
  • Disc speed (rev/h) = K × P (kW).

Instrument transformers: CT and PT

  • Kn = rated primary / rated secondary — Nominal ratio (e.g. 1000/5 = 200).
  • R = Ip / Is (CT) ; R = Vp / Vs (PT) — Actual ratio.
  • Ratio error (%) = (Kn − R) / R × 100.
  • Is ≈ Ip·Np / Ns (CT) ; Vs ≈ Vp·Ns / Np (PT) — Ideal transformation; Np, Ns = primary and secondary turns.
  • R ≈ n + (Im·sin δ + Iw·cos δ) / Is — CT actual ratio; n = Ns/Np, Im, Iw = magnetising and loss components of exciting current referred to the primary (A), δ = burden phase angle.
  • θ ≈ (Im·cos δ − Iw·sin δ) / (n·Is) — CT phase-angle error (rad).
  • P_primary = wattmeter reading × CT ratio × PT ratio — Neglecting phase-angle errors.

Potentiometers and Q-meter

  • k = V_std / l_std ; Vx = k·lx — DC potentiometer; V_std = standard-cell EMF (V), l_std, lx = balancing lengths.
  • I = V / Rs — Current measured as the drop V across a standard resistor Rs (Ω).
  • Q = ω0·L / R = 1 / (ω0·C·R) = Vc / E — Q-meter at series resonance; Vc = capacitor voltage, E = injected voltage (V).
  • f0 = 1 / (2π·√(L·(C + Cd))) — Resonant frequency including the coil's distributed capacitance Cd (F).
  • Cd = (C1 − 4·C2) / 3 — Distributed capacitance from resonance at f with C1 and at 2f with C2.
  • Q_true = Q_indicated × (C + Cd) / C — Correction for distributed capacitance.

Digital voltmeters and digital multimeters

  • Vin = Vref·N2 / N1 — Dual-slope ADC; N1 = fixed integrate count, N2 = de-integrate count.
  • T1 = N1 / fc — Integrate time; fc = clock frequency (Hz). Choose T1 = whole number of mains periods to reject hum.
  • Vo,peak = Vin·T1 / (R·C) — Integrator output at the end of T1.
  • Resolution = 1 / 10ⁿ ; S = (1 / 10ⁿ) × lowest full scale — n = number of full digits; S = DVM sensitivity (V).
  • Error = ±(a/100 × reading + b × resolution) — DVM accuracy specification ±(a % of reading + b digits).
  • t_conv = n × T_clock — Successive-approximation conversion time for n bits.

Cathode ray oscilloscope and digital storage oscilloscope

  • D = L·l·Vd / (2·d·Ea) — CRT deflection (m); L = plate centre to screen (m), l = plate length (m), d = plate spacing (m), Vd = deflecting voltage (V), Ea = accelerating voltage (V).
  • S = L·l / (2·d·Ea) ; G = 1 / S — Deflection sensitivity (m/V) and deflection factor (V/m).
  • V = divisions × V/div (× probe attenuation) ; t = divisions × time/div — Reading a display.
  • tr ≈ 0.35 / BW — Rise time (s) of a first-order system of −3 dB bandwidth BW (Hz).
  • tr,displayed = √(tr,signal² + tr,scope² + tr,probe²).
  • Rp·Cp = Rin·Cin — Compensated 10:1 probe.
  • fs > 2·fmax ; time window = record length / fs — DSO sampling (Nyquist) and capture window.

Frequency, time interval and phase measurement

  • f = 1 / T — Frequency (Hz) and period (s).
  • f = N / Tg ; ±1 count error = ±1 / (f·Tg) — Counter in frequency mode; Tg = gate time (s).
  • T = N / fc ; ±1 count error = ±f / fc — Counter in single-period mode; fc = clock frequency (Hz).
  • f_crossover = √(fc / Tg) — Frequency at which both modes give equal ±1 count error.
  • φ = 360°·Δt / T — Phase difference from time shift Δt (dual-trace or time-interval counter).
  • sin φ = y0 / y_max — Phase from a Lissajous ellipse; y0 = Y-axis intercept, y_max = maximum vertical deflection.
  • fy / fx = n_h / n_v — Lissajous frequency ratio; n_h, n_v = tangency points on horizontal and vertical edges.

True RMS meters and electronic counters

  • V_rms = √((1/T)·∫ v² dt) ; V_avg = (1/T)·∫ |v| dt — RMS and rectified average (V).
  • kf = V_rms / V_avg ; kc = V_peak / V_rms — Form factor (sine 1.111, square 1, triangle 1.155) and crest factor (sine 1.414, square 1, triangle 1.732).
  • Reading = 1.111 × V_avg — Full-wave average-responding meter calibrated for sine waves.
  • V_rms = √(V_dc² + V1² + V2² + …) — RMS of a signal with DC and harmonic components (RMS values).
  • ratio = fA / fB — Counter ratio mode (counts of A during one cycle of B).

Shielding, grounding and noise in measurements

  • Vn = √(4·k·T·R·B) — Thermal noise voltage (V RMS); k = 1.38 × 10⁻²³ J/K, T in K, R in Ω, B = bandwidth (Hz).
  • In = √(2·q·I·B) — Shot noise current (A RMS); q = 1.602 × 10⁻¹⁹ C, I = DC current (A).
  • SNR (dB) = 10·log₁₀(Ps / Pn) = 20·log₁₀(Vs / Vn).
  • CMRR = Ad / Acm ; CMRR (dB) = 20·log₁₀(Ad / Acm) — Common-mode rejection ratio of a differential amplifier.
  • Vo = Ad·(Vd + Vcm / CMRR) — Output including common-mode error; Vd = differential, Vcm = common-mode input (V).
  • Vn ≈ ω·R·C·V — Capacitive pickup (ωRC ≪ 1); C = stray capacitance (F), R = victim node resistance to ground (Ω), V = interfering voltage (V).
  • Vn = ω·M·I — Inductive pickup; M = mutual inductance (H), I = interfering current (A RMS).